Coulometric Titrators - Aquapal III
ProductSpecs
|
Standard Specs: |
Range |
1 ppm to 100% |
|
Speed |
2 mg/minute |
|
|
Sensitivity |
0.1 ug |
|
|
Precision |
0.5% at 1mg |
|
|
Drift Correction |
Automatic |
|
|
Input/Output |
Built-in RS232C for balance interface or computer |
|
|
Unique Specs: |
Single Reagent system available |
|
|
Carry case offers portability for use in the field. |
||
|
Compact size of 14"W x 8"D x 10.8"H or 350 x 200 x 280mm |
||
|
Light weight of 15 lbs. or 6.8kg |
||
|
Built in printer with standard paper backs up test with hard copy results. |
||
|
Obtain results in percent solubility. |
||
|
Simple menu operation |
||
|
Run tests by weight, volume or dilution |
||
|
Result Specs: |
Result Format |
ppm - % - ug |
|
Data Calculation |
mean, standard deviation and coeffiicient of variation |
|
|
Memory |
stores from 1 -99999 sample I.D. numbers |
|
|
Other Specs: |
Tests according to ASTM D4928, D1533, D1744 and equivalent methods. |
|
|
CE approved manufacturer. |
||