CSC Scientific Blog

A blog about test equipment

Subscribe in a reader

Subscribe via E-mail

Your email:

CSC Scientific Blog

Current Articles | RSS Feed RSS Feed

Karl Fischer Moisture – Answers to Questions We Get Asked

  
  
  
  

When People are first introduced to the Karl Fischer Moisture Determination Method, eyes glaze overKarl fischer Stirrer Bar and we can perceive a mental “Why did I Ask?

If you have any history with moisture analysis, you will have found, that for some applications, the Karl Fischer Titration Method is the best and sometimes the only way to get an accurate moisture measurement.

Most non-chemists react to the thought of titrations with “Not for me – do I have to do this?” Even the chemical formula for the Karl Fischer reactionKarl fischer Formulais mind blowing.

However, these days most of the pain is removed with automatic Karl Fischer instruments such as our Aquapal III, that make conducting Karl Fischer tests quite simple. There are still times when your not sure if the equipment is reading correctly or when it gives arcane messages like "Over Titration" that you feel panic. A host of other subtle problems bring on the question “What does this Mean?

Recently Hank Levi, a colleague of ours at Scientific Gear, developed a list of 20 questions that are periodically asked. He also developed concise answers. They are questions like;

  • Why won't my instrument get to the Ready Mode?
  • What kind of reagent should I use?
  • Is my instrument giving me correct results?
  • How much sample should I use?

We would like to share these with you. If you click on the button you can down-load the full list of  Karl Fisher questions and answers.


20-karl-fischer-questions-answers

Hank tells us that some of his customers post this list in the lab where it is available for reference at any time to the staff conducting Karl Fischer moisture determinations.

Maybe this will help reduce trepidation about Karl Fischer and help get good repeatable results.

We hope that you check out this list and that it is useful to you. Again thanks for visiting.

I remain a wary respectful friend of Karl Fischer,

Art

P. S. Subscribe to our blog.

Comments

I am on the way to buy the sinar 6060 fixed calibratio analyzer,I have seen the 20 questions and answers for the karl fischer moisture measurements its so useful. 
Now is there questions and answers for the sinar 6060 fixed calibration analyzer? 
Posted @ Thursday, August 04, 2011 3:26 AM by Shabani
Thanks for you comment. We Don't have one yet. When we do we will advise here. 
 
Art
Posted @ Thursday, August 04, 2011 1:21 PM by Art
In our organization we do K-F on transformer oils using ASTM D-1533 on an Aquapal III. As I was learning the process it became evident that precision was lousy & no one had any clue about the accuracy of the analysis. We brought all of our Aquapals up to date with fritless generators using the single cell. The coulometric solution has been modified with 20% toluene to make the oil disperse better. A star shaped stir bar spins faster than the rod versions & swishes the oil into smaller globules. For added protection of the generator cell I installed a spacer ring on its neck so it can't be forced too far down into the cell where it can be hit by the stir bar. I did some glass blowing to modify test tubes into extra long dessicant holders for extra drying power. We use gas tight syringes with Teflon tipped plungers that are stored in horizontally held tubes with dessicant & plugged with a massive rubber cork when not in use. Every morning accuracy is checked by injecting a commercially prepared water standard in xylenes. The solutions are factory sealed in ampules. The Aquapal is run on a 45 second delay to give the stirring solution time to extract all the water from the small oil droplets as they form during their slow inject time spanning 1/2 minute. Now we are sure of our accuracy & precision has improved remarkably.
Posted @ Monday, August 08, 2011 2:10 PM by Andreas Mueller
Post Comment
Name
 *
Email
 *